USER GUIDE
Flow Meter Testing
Version 1 Flow Meter User Guide
5
FLOW METER TEST POINTS
The following table describes the flow meter readings for a new, rebuilt, or dirty
air pump in liters/minute. Perform all of the tests outlined in this manual and fill
out your results in the data tables on page 16. When you have completed all of the
tests, compare your results with the following table. If the flow meter reading is
equal to or lower than the dirty air pump reading, follow the steps outlined in the
action required column of the table. Pump head rebuild kits (P/N 79190003) are
available from Dynojet. Refer to the Air Fuel Ratio Module Installation and User
Guide (P/N 98295110) for rebuild instructions.
Note: If at any test point, the ball in the flow meter bounces erratically
(+/- 20L/min.) the pump heads need to be rebuilt. Pump head rebuild kits
(P/N 79190003) are available from Dynojet. Refer to the Air Fuel Ratio Module
Installation Guide (P/N 98295110) for rebuild instructions.
test point
new or rebuilt
pump dirty pump action required
test point one 35 25 or less rebuild pump heads
test point two 28 25 or less clean or replace filter
test point two—filter
assembly
less than 2 greater than 2 clean and tighten filter
housing
test point three 28 25 or less rebuild pump heads
test point three—sensor
block assembly
less than 2 greater than 2 clean and tighten sensor
test point four 30 25 or less rebuild pump heads
test point five 30 25 or less rebuild pump heads